Development of a dynamic motion platform with two independent drive systems for radiotherapy.


ABSTRACT:

Background

There are some motion platforms for radiotherapy quality assurance. However, no platform with two drive systems that can move along three axes is available.

Purpose

The purpose of this study is to develop a dynamic motion platform with two drive systems capable of three-axis motion and to evaluate its motion performance.

Methods

The developed moving platform had two drive systems that use the same equipment. Each axis of the platform used can support a maximum load of 10 kg. The motors for moving the platform in each direction are capable of a drive stroke up to 40 mm. The drive speed is 30 mm/s at maximum load fluctuation. To evaluate the static positional accuracy of this system with an arbitrary input movement, the XYZ position of each axis was measured

SUBMITTER: Saito M 

PROVIDER: S-EPMC10161022 | biostudies-literature | 2023 May