Hematoma-induced Twiddler-like phenomenon as a presentation of DBS hardware failure: Case report.


ABSTRACT: Deep brain stimulators (DBS) may fail for a multitude of reasons. We present a 79-year-old Parkinson's disease patient who suffered a DBS failure following impulse generator (IPG) replacement surgery due to the IPG flipping within an expanded capsular pocket. This creation of the pocket was unintentional, and the pocket formed around an undiagnosed postoperative hemorrhage. The syndrome could be considered "Twiddler-like" because it resulted in device flipping. There were, however, many characteristic differences between our case and classical Twiddler's syndrome. DBS neurostimulator failure due to hematoma induced device flipping should be suspected when device interrogation is impossible or there are abnormally high impedances across multiple DBS lead contacts. A plain film X-ray series