A new full-field XRF imaging station at Synchrotron Light Research Institute.
Ontology highlight
ABSTRACT: A full-field X-ray fluorescence imaging (FXI) station was recently developed at beamline BL8 of Synchrotron Light Research Institute (SLRI), Thailand. An unfocused, synchrotron X-ray beam from the bending magnet with a size of 2 mm (vertical) × 13 mm (horizontal) and photon energy of 10 keV was employed in the FXI experiments. A sample stage was tilted by 7.5° to enlarge the vertical beam size. X-ray fluorescence images were recorded by an energy-dispersive, 256 × 256 array, pn-type charge coupled device detector equipped with a polycapillary optics, providing a full-frame image size of 12.3 mm × 12.3 mm. The incident photon flux per pixel was 3 × 104 photons s-1 (100 mA)-1 and the experimental spatial resolution was 68 µm. Image processing was carried out
SUBMITTER: Klysubun W
PROVIDER: S-EPMC10161892 | biostudies-literature | 2023 May
REPOSITORIES: biostudies-literature