Effect of crown stiffness and prosthetic screw absence on the stress distribution in implant-supported restoration: A 3D finite element analysis.
Ontology highlight
ABSTRACT: This in-silico investigation evaluated the mechanical impact of Morse tape implant-abutment interface and retention system (with and without screw) and restorative materials (composite block and monolithic zirconia) by means of a three-dimensional finite element analysis (3D-FEA). Four 3D models were designed for the lower first molar. A dental implant (4.5 × 10 mm B&B Dental Implant Company) was digitized (micro CT) and exported to computer-aided design (CAD) software. Non-uniform rational B-spline surfaces were reconstructed, generating a 3D volumetric model. Four different models were generated with the same Morse-type connection, but with a different locking system (with and without active screw) and a different crown material made of composite block and zirconia. The D2 bone type, whi