Unknown

Dataset Information

5T magnetic resonance imaging: radio frequency hardware and initial brain imaging.


ABSTRACT:

Background

We aimed to demonstrate the feasibility of generating high-resolution human brain magnetic resonance imaging (MRI) at 5 Tesla (T) using a quadrature birdcage transmit/48-channel receiver coil assembly.

Methods

A quadrature birdcage transmit/48-channel receiver coil assembly was designed for human brain imaging at 5T. The radio frequency (RF) coil assembly was validated by electromagnetic (EM) simulations and phantom imaging experimental studies. The simulated B1+ field inside a human head phantom and inside a human head model generated by the birdcage coils driven in circularly polarized (CP) mode at 3T, 5T and 7T was compared. Signal-to-noise ratio (SNR) maps, the inverse g-factor maps for evaluation of parallel imaging performance, anatomic images, angiography i

SUBMITTER: Wei Z 

PROVIDER: S-EPMC10167427 | biostudies-literature | 2023 May

REPOSITORIES: biostudies-literature