Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients.


ABSTRACT:

Background

An accurate SpO2 value is critical in order to optimally titrate oxygen delivery to patients and to follow oxygenation guidelines. Limited prospective data exist on real-world performance of pulse oximeters in critically ill patients. The objective of this study was to assess accuracy and bias of the SpO2 values measured by several oximeters in hospitalized subjects.

Methods

We included stable adults in the ICU with an arterial catheter in place. Main exclusion criteria were poor SpO2 signal and SpO2 > 96%. In each subject, we simultaneously evaluated 4 oximeters: Nonin (Plymouth, Minnesota) embedded in the FreeO2 device (OxyNov, Québec City, Québec, Canada), Masimo

SUBMITTER: Blanchet MA 

PROVIDER: S-EPMC10171338 | biostudies-literature | 2023 May