Breath-by-breath measurement of exhaled ammonia by acetone-modifier positive photoionization ion mobility spectrometry via online dilution and purging sampling.
Ontology highlight
ABSTRACT: Exhaled ammonia (NH3) is an essential noninvasive biomarker for disease diagnosis. In this study, an acetone-modifier positive photoionization ion mobility spectrometry (AM-PIMS) method was developed for accurate qualitative and quantitative analysis of exhaled NH3 with high selectivity and sensitivity. Acetone was introduced into the drift tube along with the drift gas as a modifier, and the characteristic NH3 product ion peak of (C3H6O)4NH4+ (K0 = 1.45 cm2/V·s) was obtained through the ion-molecule reaction with acetone reactant ions (C3H6O)2H+ (K0 = 1.87 cm2/V·s), which significantly increased the pe