Visual Performance of Eyes with Residual Refractive Errors after Implantation of an Extended Vision Intraocular Lens.


ABSTRACT:

Background

To analyze the tolerance on distance vision of different combined residual astigmatic situations in patients implanted with a novel wavefront shaping extended depth of focus (EDoF) intraocular lens (IOL).

Methods

The study included patients implanted with the Acrysof® IQ Vivity® IOL. Uncorrected (UDVA) and corrected distance visual acuity (CDVA) were measured three months after surgery, considering CDVA as the reference situation of the study. Distance VA was also measured in different refractive situations: (A) with 0.50 diopters (D) of positive (myopization) and negative (hyperopization) defocus and (B) with a residual mixed astigmatic refraction induced by adding a combination of -0.25 D spherical and 0.50 D cylindrical lenses placed in vertical (against the ru