The accuracy and precision of acetabular implant measurements from CT imaging.


ABSTRACT: The placement of acetabular implant components determines the short- and long-term outcomes of total hip replacement (THR) and a number of tools have been developed to assist the surgeon in achieving cup orientation to match the surgical plan. However, the accuracy and precision of 3D-CT for the measurement of acetabular component position and orientation is yet to be established. To investigate this, we compared measurements of cobalt chrome acetabular components implanted into 2 different bony pelvic models between a coordinate measuring Faro arm and 3 different low dose CT images, including 3D-CT, 2D anterior pelvic plane (APP) referenced CT and 2D scanner referenced (SR) CT. Intra-observer differences were assessed using the Intraclass correlation coefficient (ICC). The effect of imaging the pelvis positioned in 3 different orientations within the CT scanner was also assessed. The measured parameters were the angles of inclination and version. 3D-CT measurements were found to closely match the "true values" of the component position measurements, compared with the 2D-CT methods. ICC analysis also showed good agreement between the coordinate measuring arm (CMA) and 3D-CT but poor agreement between the 2D SR method, in the results from two observers. When using the coordinate system of the CT scanner, the measurements consistently produced the greatest error; this method yielded values up to 34° different from the reference digitising arm. However, the difference between the true inclination and version angles and those measured from 3D APP CT was below half a degree in all cases. We concluded that low radiation dose 3D-CT is a validated reference standard for the measurement of acetabular cup orientation.

SUBMITTER: Henckel J 

PROVIDER: S-EPMC10175694 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature

altmetric image

Publications

The accuracy and precision of acetabular implant measurements from CT imaging.

Henckel Johann J   Ramesh Angelika A   Hothi Harry H   Richards Robin R   Di Laura Anna A   Hart Alister A  

Frontiers in bioengineering and biotechnology 20230428


The placement of acetabular implant components determines the short- and long-term outcomes of total hip replacement (THR) and a number of tools have been developed to assist the surgeon in achieving cup orientation to match the surgical plan. However, the accuracy and precision of 3D-CT for the measurement of acetabular component position and orientation is yet to be established. To investigate this, we compared measurements of cobalt chrome acetabular components implanted into 2 different bony  ...[more]

Similar Datasets

| S-EPMC8381926 | biostudies-literature
| S-EPMC10544156 | biostudies-literature
| S-EPMC3026088 | biostudies-literature
| S-EPMC3353415 | biostudies-literature
| S-EPMC8259187 | biostudies-literature
| S-EPMC8376669 | biostudies-literature
| S-EPMC6121742 | biostudies-literature
| S-EPMC11465135 | biostudies-literature
| S-EPMC11348571 | biostudies-literature
| S-EPMC2858813 | biostudies-other