Unknown

Dataset Information

The accuracy and precision of acetabular implant measurements from CT imaging.


ABSTRACT: The placement of acetabular implant components determines the short- and long-term outcomes of total hip replacement (THR) and a number of tools have been developed to assist the surgeon in achieving cup orientation to match the surgical plan. However, the accuracy and precision of 3D-CT for the measurement of acetabular component position and orientation is yet to be established. To investigate this, we compared measurements of cobalt chrome acetabular components implanted into 2 different bony pelvic models between a coordinate measuring Faro arm and 3 different low dose CT images, including 3D-CT, 2D anterior pelvic plane (APP) referenced CT and 2D scanner referenced (SR) CT. Intra-observer differences were assessed using the Intraclass correlation coefficient (ICC). The effect of imagi

SUBMITTER: Henckel J 

PROVIDER: S-EPMC10175694 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature