Slice encoding for the reduction of outflow signal artifacts in cine balanced SSFP imaging.


ABSTRACT:

Purpose

Standard balanced SSFP (bSSFP) cine MRI often suffers from blood outflow artifacts. We propose a method that spatially encodes these outflowing spins to reduce their effects in the intended slice.

Methods

Bloch simulations were performed to characterize through-plane flow and to investigate how the use of phase encoding along the slice select's direction ("slice encoding") could alleviate its issues. Phantom scans and in vivo cines were acquired on a 3T system, comparing the standard 2D acquisition to the proposed slice-encoding method. Nineteen healthy volunteers were recruited for short-axis and horizontal long-axis oriented scans. An expert radiologist evaluated each slice-encoded/standard cine pairs in a rank comparison test and graded their quality on a 1-5 scal