Use of adhesive cranial bone flap fixation without hardware to improve mechanical strength, resist cerebrospinal fluid leakage, and maintain anatomical alignment: a laboratory study.


ABSTRACT:

Objective

Titanium plates and screws (TPS) are the current standard of care for fixation of cranial bone flaps. These materials have been used for decades but have known potential complications, including flap migration, bone resorption/incomplete osseous union, hardware protrusion, cosmetic deformity, wound infection/dehiscence, and cerebrospinal fluid (CSF) leakage. This study evaluated the efficacy of a novel mineral-organic bone adhesive (Tetranite) for cranial bone flap fixation.

Methods

Craniotomy bone flaps created in human cadaveric skulls were tested under quasistatic and impact loading in the following conditions: 1) uncut skull; 2) bone flaps fixated with TPS alone; and 3) bone flaps fixated with bone adhesive alone. All fixative surgical procedures were performed

SUBMITTER: Smith TR 

PROVIDER: S-EPMC10193477 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature