Radiation Exposure Reduction by Digital Variance Angiography in Lower Limb Angiography: A Randomized Controlled Trial.


ABSTRACT:

Background

digital variance angiography (DVA) provides higher image quality than digital subtraction angiography (DSA). This study investigates whether the quality reserve of DVA allows for radiation dose reduction during lower limb angiography (LLA), and compares the performance of two DVA algorithms.

Methods

this prospective block-randomized controlled study enrolled 114 peripheral arterial disease patients undergoing LLA into normal dose (ND, 1.2 µGy/frame, n = 57) or low-dose (LD, 0.36 µGy/frame, n = 57) groups. DSA images were generated in both groups, DVA1 and DVA2 images were generated in the LD group. Total and DSA-related radiation dose area product (DAP) were analyzed. Image quality was assessed on a 5-grade Likert scale by six readers.

Results

the total and DSA-related DAP were reduced by 38% and 61% in the LD group. The overall visual evaluation scores (median (IQR)) of LD-DSA (3.50 (1.17)) were significantly lower than the ND-DSA scores (3.83 (1.00), p < 0.001). There was no difference between ND-DSA and LD-DVA1 (3.83 (1.17)), but the LD-DVA2 scores were significantly higher (4.00 (0.83), p < 0.01). The difference between LD-DVA2 and LD-DVA1 was also significant (p < 0.001).

Conclusions

DVA significantly reduced the total and DSA-related radiation dose in LLA, without affecting the image quality. LD-DVA2 images outperformed LD-DVA1, therefore DVA2 might be especially beneficial in lower limb interventions.

SUBMITTER: Sotonyi P 

PROVIDER: S-EPMC10219428 | biostudies-literature | 2023 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Radiation Exposure Reduction by Digital Variance Angiography in Lower Limb Angiography: A Randomized Controlled Trial.

Sótonyi Péter P   Berczeli Márton M   Gyánó Marcell M   Legeza Péter P   Mihály Zsuzsanna Z   Csobay-Novák Csaba C   Pataki Ákos Á   Juhász Viktória V   Góg István I   Szigeti Krisztián K   Osváth Szabolcs S   Kiss János P JP   Nemes Balázs B  

Journal of cardiovascular development and disease 20230430 5


<h4>Background</h4>digital variance angiography (DVA) provides higher image quality than digital subtraction angiography (DSA). This study investigates whether the quality reserve of DVA allows for radiation dose reduction during lower limb angiography (LLA), and compares the performance of two DVA algorithms.<h4>Methods</h4>this prospective block-randomized controlled study enrolled 114 peripheral arterial disease patients undergoing LLA into normal dose (ND, 1.2 µGy/frame, <i>n</i> = 57) or lo  ...[more]

Similar Datasets

| S-EPMC8575921 | biostudies-literature
| S-EPMC10384110 | biostudies-literature
| S-EPMC7369257 | biostudies-literature
| S-EPMC10415441 | biostudies-literature
| S-EPMC7683322 | biostudies-literature
| S-EPMC8556952 | biostudies-literature
2020-10-22 | GSE159610 | GEO
| S-EPMC4405855 | biostudies-literature
| S-EPMC5117579 | biostudies-literature
| S-EPMC8325805 | biostudies-literature