Application of quality-by-design for adopting an environmentally green fluorogenic determination of benoxinate hydrochloride in eye drops and artificial aqueous humour.
Ontology highlight
ABSTRACT: Herein, a sensitive and selective spectrofluorimetric method has been developed for the determination of the ocular local anesthetic benoxinate hydrochloride (BEN-HCl) in eye drops and artificial aqueous humour. The proposed method is based on the interaction of fluorescamine with the primary amino group of BEN-HCl at room temperature. Following the excitation of the reaction product at 393 nm, the emitted relative fluorescence intensity (RFI) was measured at 483 nm. The key experimental parameters were carefully examined and optimized by adopting an analytical quality-by-design approach. The method used a two-level full factorial design (24 FFD) to obtain the optimum RFI of the reaction product. The calibration curve was linear at the range of 0.10-1.0 μg/mL of BEN-HCl with sen
SUBMITTER: El Hamd MA
PROVIDER: S-EPMC10219972 | biostudies-literature | 2023 May
REPOSITORIES: biostudies-literature
ACCESS DATA