ABSTRACT: Objective
To evaluate the sensitivity and specificity of the ABCD progression display for keratoconus progression.Methods
Data was collected from patients that underwent at least two Pentacam assessments 6 months apart. Sensitivity and specificity were calculated for the ABCD progression display. Progression was defined by criterion 1: change in two ABCD parameters above 80% confidence interval (CI) or criterion 2: change in one ABCD parameter above 95%CI. Receiver operating characteristic analysis compared the area under the curve (AUC) of all ABCD parameter combinations.Results
Thirty eyes were evaluated over a median time of 10.3 months. Progression by criterion 1 resulted in a sensitivity of 61.9% and specificity of 88.9%. Progression by criterion 2 resulted in