Accuracy of a Smartwatch to Assess Heart Rate Monitoring and Atrial Fibrillation in Stroke Patients.
Ontology highlight
ABSTRACT: (1) Background: Consumer smartwatches may be a helpful tool to screen for atrial fibrillation (AF). However, validation studies on older stroke patients remain scarce. The aim of this pilot study from RCT NCT05565781 was to validate the resting heart rate (HR) measurement and the irregular rhythm notification (IRN) feature in stroke patients in sinus rhythm (SR) and AF. (2) Methods: Resting clinical HR measurements (every 5 min) were assessed using continuous bedside ECG monitoring (CEM) and the Fitbit Charge 5 (FC5). IRNs were gathered after at least 4 h of CEM. Lin's concordance correlation coefficient (CCC), Bland-Altman analysis, and mean absolute percentage error (MAPE) were used for agreement and accuracy assessment. (3) Results: In all, 526 individual pairs of measurements were obta