Improving the Accuracy of Pdot-Based Continuous Glucose Monitoring by Using External Ratiometric Calibration.


ABSTRACT: Continuous glucose monitoring (CGM) allows type I and II diabetes patients to track changes in their glucose levels, allowing detection of impending hypoglycemia or hyperglycemia. Polymer dots (Pdots) are candidates for use in implanted CGM systems due to their exceptional brightness, photostability, sensitivity, and biocompatibility. However, Pdot glucose transducers are oxygen-dependent, and changes in tissue oxygen levels affect their measurement accuracy. Here, we describe an external ratiometric calibration method that corrects for changes in tissue oxygen levels to improve measurement accuracy. This method uses the ratio of oxygen concentrations inside and outside the Pdot glucose transducer as an indicator of glucose concentration to correct for signal deviations caused by tissue ox