Hemostatic Complications During Neonatal Extracorporeal Membrane Oxygenation: Roller Pump and Centrifugal Pump Driven Circuits.


ABSTRACT: Recently three different neonatal extracorporeal membrane oxygenation (ECMO) circuits have been employed in our clinic. These circuits were compared for clotting and bleeding complications. Initially, we used an ECMO circuit containing a roller pump and venous bladder without severe complications. Manufacturing of circuit components was discontinued, necessitating the replacement of this circuit by a circuit with a centrifugal pump with 3/8 inch inlet and outlet. Acute increase of oxygenator resistance requiring emergency changeout became unexpectedly a regularly occurring complication. The increase in resistance was suspected to be caused by oxygenator clotting, although oxygenator function was preserved. To prevent this complication, we changed to a levitating centrifugal pump with 1/4 i