ABSTRACT: Purpose
The magnetization transfer and relaxation normalized amide proton transfer (MRAPT) analysis is promising to provide a highly pH-specific mapping of tissue acidosis, complementing commonly used CEST asymmetry analysis. We aimed to develop a fast B0 inhomogeneity correction algorithm for acute stroke magnetization transfer and relaxation normalized amide proton transfer imaging without Z-spectral interpolation.Methods
The proposed fast field inhomogeneity correction describes B0 artifacts with linear regression. We compared the new algorithm with the routine interpolation correction approach in CEST imaging of a dual-pH phantom. The fast B0 correction was further evaluated in amide proton transfer imaging of normal and acute stroke rat