Unknown

Dataset Information

Diagnostic performance of four lateral flow immunoassays for COVID-19 antibodies in Peruvian population.


ABSTRACT: Serological assays have been used in seroprevalence studies to inform the dynamics of COVID-19. Lateral flow immunoassay (LFIA) tests are a very practical technology to use for this objective; however, one of their challenges may be variable diagnostic performance. Given the numerous available LFIA tests, evaluation of their accuracy is critical before real-world implementation. We performed a retrospective diagnostic evaluation study to independently determine the diagnostic accuracy of 4 different antibody-detection LFIA tests: Now Check (Bionote), CareStart (Access bio), Covid-19 BSS (Biosynex) and OnSite (CTK Biotech). The sample panel was comprised of specimens collected and stored in biobanks; specifically, specimens that were RT-PCR positive for SARS-CoV-2 collected at various times

SUBMITTER: Calderon-Flores R 

PROVIDER: S-EPMC10237462 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature