Built-in wavelet-induced smoothness to reduce plan complexity in intensity modulated radiation therapy (IMRT).
Ontology highlight
ABSTRACT: Objective.Reducing plan complexity in intensity modulated radiation therapy (IMRT) to ensure dosimetric accuracy and delivery efficiency of the radiation treatment plans. We propose a novel approach by representing the beamlet intensities using an incomplete wavelet basis that explicitly excludes fluctuating intensity maps from the decision space (explicit hard constraint). This technique provides a built-in wavelet-induced smoothness that improves both dosimetric plan quality and delivery efficiency.Approach.The beamlet intensity maps need to be especially smooth in the leaf travel direction (referred to as theX-direction). We treat the intensity map of each beam as a 2D image and represent it using the wavelets corresponding to low-frequency changes in theX
SUBMITTER: Tefagh M
PROVIDER: S-EPMC10243219 | biostudies-literature | 2023 Mar
REPOSITORIES: biostudies-literature
ACCESS DATA