Design of an adaptable intrafascicular electrode (AIR) for selective nerve stimulation by model-based optimization.
Ontology highlight
ABSTRACT: Peripheral nerve stimulation is being investigated as a therapeutic tool in several clinical scenarios. However, the adopted devices have restricted ability to obtain desired outcomes with tolerable off-target effects. Recent promising solutions are not yet employed in clinical practice due to complex required surgeries, lack of long-term stability, and implant invasiveness. Here, we aimed to design a neural interface to address these issues, specifically dimensioned for pudendal and sacral nerves to potentially target sexual, bladder, or bowel dysfunctions. We designed the adaptable intrafascicular radial electrode (AIR) through realistic computational models. They account for detailed human anatomy, inhomogeneous anisotropic conductance, following the trajectories of axons along curving
SUBMITTER: Ciotti F
PROVIDER: S-EPMC10246853 | biostudies-literature | 2023 May
REPOSITORIES: biostudies-literature
ACCESS DATA