Measurement of serum anti-Mullerian hormone by revised Gen II or automated assay: Reproducibility under various blood/serum storage conditions.


ABSTRACT:

Objective

We investigated the agreement between anti-Müllerian hormone (AMH) levels measured with revised Gen II (rev-Gen II) and automated AMH (Access) assays and evaluated the reproducibility of each method under various blood/serum storage conditions.

Methods

AMH levels in blood samples from 74 volunteers were measured by rev-Gen II and Access assays under various conditions: immediate serum separation and AMH measurement (fresh control); serum stored at -20 °C and AMH measured after 48 hours, 1 week, and 2 years; serum stored at 0 to 4 °C and AMH measured after 48 hours and 1 week; and blood kept at room temperature and delayed serum separation after 48 hours and 1 week, with immediate AMH measurement.

Results

In fresh controls, all rev-Gen II-AMH values were high