ABSTRACT: Hypothesis
High intracochlear pressure transients associated with cochlear implant placement are reduced with smaller, non-styleted arrays, and longer insertion durations.Background
With increasing focus on hearing preservation during cochlear implant surgery, atraumatic technique is of the utmost importance. Previous studies revealed that high intensity pressure transients can be generated during the insertion of implant electrodes. Resulting acoustic trauma may be one contributing factor to postoperative loss of residual hearing.Methods
Thirty ears in cadaveric specimens were surgically prepared with placement of intracochlear pressure sensors. Sequential implant insertions were made over 10, 30, or 60 seconds using seven randomly ordered electrode styles. Pressur