Reducing contrast-agent volume and radiation dose in CT with 90-kVp tube voltage, high tube current modulation, and advanced iteration algorithm.
Ontology highlight
ABSTRACT: Increasing utilization of computed tomography (CT) has raised concerns regarding CT radiation dose and technology has been developed to achieve an appropriate balance between image quality, radiation dose, and the amount of contrast material. This study was planned to evaluate the image quality and radiation dose in pancreatic dynamic computed tomography (PDCT) with 90-kVp tube voltage and reduction of the standard amount of contrast agent, compared with 100-kVp PDCT of the research hospital's convention. Total of 51 patients with both CT protocols were included. The average Hounsfield units (HU) values of the abdominal organs and image noise were measured for objective image quality analysis. Two radiologists evaluated five categories of image qualities such as subjective image noise, vis