More than algorithms: an analysis of safety events involving ML-enabled medical devices reported to the FDA.


ABSTRACT:

Objective

To examine the real-world safety problems involving machine learning (ML)-enabled medical devices.

Materials and methods

We analyzed 266 safety events involving approved ML medical devices reported to the US FDA's MAUDE program between 2015 and October 2021. Events were reviewed against an existing framework for safety problems with Health IT to identify whether a reported problem was due to the ML device (device problem) or its use, and key contributors to the problem. Consequences of events were also classified.

Results

Events described hazards with potential to harm (66%), actual harm (16%), consequences for healthcare delivery (9%), near misses that would have led to harm if not for intervention (4%), no harm or consequences (3%), and complaints (2%). Wh