Investigation of the Influence of Pulse Duration and Application Mode on Microsecond Laser Microsurgery of the Retinal Pigment Epithelium.
Ontology highlight
ABSTRACT: Optical microsurgery confined to the retinal pigment epithelium (RPE) requires locally optimized laser parameters and reliable real-time feedback dosimetry (RFD) to prevent unwanted neuroretinal overexposure. This study aimed to compare pulses of different durations and application modes (single, ramp, burst). Moreover, optical coherence tomography (OCT)-based RFD was investigated in an ex vivo experiment, utilizing nine porcine eyes that were exposed to laser pulses of 8, 12, 16 and 20 µs duration (wavelength: 532 nm, exposure area: 90 × 90 µm2, radiant exposure: 247 to 1975 mJ/µm2). Simultaneously, time-resolved OCT M-scans were recorded (central wavelength: 870 nm, scan rate: 85 kHz) for RFD. Post irradiation, retinal changes were assessed with color fundus photogr
SUBMITTER: Burri C
PROVIDER: S-EPMC10302843 | biostudies-literature | 2023 Jun
REPOSITORIES: biostudies-literature
ACCESS DATA