Ontology highlight
ABSTRACT: Objectives
To investigate discordance in oxy-hemoglobin saturation measured both by pulse oximetry (SpO2) and arterial blood gas (ABG, SaO2) among critically ill coronavirus disease 2019 (COVID-19(+)) patients compared to COVID-19(-) patients.Methods
Paired SpO2 and SaO2 readings were collected retrospectively from consecutive adult admissions to four critical care units in the United States between March and May 2020. The primary outcome was the rate of discordance (|SaO2-SpO2|>4%) in COVID-19(+) versus COVID-19(-) patients. The odds each cohort could have been incorrectly categorized as having a PaO2/FiO2 above or below 150 by their SpO2: Fractional inhaled oxygen ratio (pul
SUBMITTER: Fox TH
PROVIDER: S-EPMC10331117 | biostudies-literature | 2023 Jul
REPOSITORIES: biostudies-literature