An integrated and fast imaging quality assurance phantom for a 0.35 T magnetic resonance imaging linear accelerator.


ABSTRACT:

Purpose

Periodic imaging quality assurance (QA) of magnetic resonance imaging linear accelerator (MRL) is critical. The feasibility of a new MRL imaging phantom used for QA in the low field was evaluated with automated image analysis of various parameters for accuracy and reproducibility.

Methods and materials

The new MRL imaging phantom was scanned across every 30 degrees of the gantry, having the on/off state of the linac in a low-field MRL system using three magnetic resonance imaging sequences: true fast imaging with steady-state precession (TrueFISP), T1 weighted (T1W), and T2 weighted (T2W). The DICOM files were used to calculate the imaging parameters: geometric distortion, uniformity, resolution, signal-to-noise ratio (SNR), and laser alignment. The point spread func

SUBMITTER: Sohn JJ 