Design and implementation of high sampling rate and multichannel wireless recorder for EEG monitoring and SSVEP response detection.


ABSTRACT:

Introduction

The collection and process of human brain activity signals play an essential role in developing brain-computer interface (BCI) systems. A portable electroencephalogram (EEG) device has become an important tool for monitoring brain activity and diagnosing mental diseases. However, the miniaturization, portability, and scalability of EEG recorder are the current bottleneck in the research and application of BCI.

Methods

For scalp EEG and other applications, the current study designs a 32-channel EEG recorder with a sampling rate up to 30 kHz and 16-bit accuracy, which can meet both the demands of scalp and intracranial EEG signal recording. A fully integrated electrophysiology microchip RHS2116 controlled by FPGA is employed to build the EEG recorder, and the desi

SUBMITTER: Li R 

PROVIDER: S-EPMC10339741 | biostudies-literature | 2023