Ontology highlight
ABSTRACT: Objective
Processed electroencephalography (pEEG) is used to monitor depth-of-anesthesia during cardiopulmonary bypass (CPB). The SedLine device has been recently introduced for pEEG monitoring. However, the effect of hypothermia on its parameters during CPB is unknown. Accordingly, we aimed to investigate temperature-induced changes in SedLine-derived pEEG parameters during CPB.Design
Prospective observational study.Setting
Cardiac surgery operating theatre.Participants
28 patients undergoing elective cardiac surgery with CPB.Interventions
We continuously measured patient state index (PSI), suppression ratio (SR), bilateral spectral edge frequency (SEF) and temperature. We used linear mixed modelling with fixed and random effects to study the intera
SUBMITTER: Belletti A
PROVIDER: S-EPMC10352607 | biostudies-literature | 2023
REPOSITORIES: biostudies-literature