Phospholipid Removal for Enhanced Chemical Exposomics in Human Plasma.
Ontology highlight
ABSTRACT: The challenge of chemical exposomics in human plasma is the 1000-fold concentration gap between endogenous substances and environmental pollutants. Phospholipids are the major endogenous small molecules in plasma, thus we validated a chemical exposomics protocol with an optimized phospholipid-removal step prior to targeted and non-targeted liquid chromatography high-resolution mass spectrometry. Increased injection volume with negligible matrix effect permitted sensitive multiclass targeted analysis of 77 priority analytes; median MLOQ = 0.05 ng/mL for 200 μL plasma. In non-targeted acquisition, mean total signal intensities of non-phospholipids were enhanced 6-fold in positive (max 28-fold) and 4-fold in negative mode (max 58-fold) compared to a control method without phospholipid removal
SUBMITTER: Sdougkou K
PROVIDER: S-EPMC10357568 | biostudies-literature | 2023 Jul
REPOSITORIES: biostudies-literature
ACCESS DATA