Development of non-electrically controlled SalivaDirect LAMP (NEC-SD-LAMP), a new nonelectrical infectious disease testing method.


ABSTRACT: In this study, non-electrically controlled SalivaDirect loop-mediated isothermal amplification (NEC-SD-LAMP), which can detect infections by amplifying viral DNA expression in saliva without using electrical control systems, was developed. By this method, only by adding water to the device, viral DNA was extracted from saliva using SalivaDirect, the extracted DNA was amplified via loop-mediated isothermal amplification (LAMP), and the results were visually confirmed. Melting palmitic acid maintained the optimal temperature for the LAMP reaction, as the temperature of palmitic acid is maintained at 62.9 °C, its melting point. By exploiting the proximity of the melting point to the optimal temperature for LAMP, LAMP can be performed without electricity. We detected several viruses in the sal

SUBMITTER: Kimura Y 

PROVIDER: S-EPMC10362045 | biostudies-literature | 2023 Jul