Comparison of onset of neuromuscular blockade with electromyographic and acceleromyographic monitoring: a prospective clinical trial.


ABSTRACT:

Background

Reliable devices that quantitatively monitor the level of neuromuscular blockade after neuromuscular blocking agents' administration are crucial. Electromyography and acceleromyography are two monitoring modalities commonly used in clinical practice. The primary outcome of this study is to compare the onset of neuromuscular blockade, defined as a Train-Of-Four Count (TOFC) equal to 0, as measured by an electromyography-based device (TetraGraph) and an acceleromyography-based device (TOFscan). The secondary outcome was to compare intubating conditions when one of these two devices reached a TOFC equal to 0.

Methods

One hundred adult patients scheduled for elective surgery requiring neuromuscular blockade were enrolled. Prior to induction of anesthesia, TetraGraph e

SUBMITTER: Chaves-Cardona HE 

PROVIDER: S-EPMC10362458 | biostudies-literature | 2023 Jul-Aug