SALAD: Syringe-based Arduino-operated Low-cost Antibody Dispenser.
Ontology highlight
ABSTRACT: Lateral Flow Assays (LFA) have been one of the most widely adopted technologies in clinical diagnosis over recent years, especially during the COVID-19 pandemic, due to their feasibility, compactness, and rapid readout. However, the precise dispensing of antibodies-a key part of the fabrication process-requires costly line dispenser equipment, which poses a challenge to researchers with limited budgets. This study aims to resolve this key issue by introducing a Syringe-based Arduino-operated Low-cost Antibody Dispenser (SALAD). By utilizing a microneedle, stepper motor-driven syringe pump, and conveyor belt, SALAD can form micro-droplets to create an even band of antibodies. Our evaluation results showed comparable performance between SALAD and a commercialized model - Claremont ALFRD, wit
SUBMITTER: Le APH
PROVIDER: S-EPMC10366588 | biostudies-literature | 2023 Sep
REPOSITORIES: biostudies-literature
ACCESS DATA