Ontology highlight
ABSTRACT:
SUBMITTER: Patel M
PROVIDER: S-EPMC10368176 | biostudies-literature | 2023 Jun
REPOSITORIES: biostudies-literature
Patel Mehulkumar M Patel Mehulkumar M Jamiolkowski Megan A MA Vejendla Anjali A Bentley Viviana V Malinauskas Richard A RA Lu Qijin Q
ASAIO journal (American Society for Artificial Internal Organs : 1992) 20230227 6
To develop and standardize a reliable in vitro dynamic thrombogenicity test protocol, the key test parameters that could impact thrombus formation need to be investigated and understood. In this study, we evaluated the effect of temperature on the thrombogenic responses (thrombus surface coverage, thrombus weight, and platelet count reduction) of various materials using an in vitro blood flow loop test system. Whole blood from live sheep and cow donors was used to assess four materials with vary ...[more]