Successful Retrieval of the Distal Anti-embolic Device after Carotid Stenting Using the Balloon Bridge Technique.


ABSTRACT:

Background

During carotid artery stenting (CAS), retrieval of the distal umbrella portion of the anti-embolic device (AED) could be difficult. Herein, we report a case of left CAS managed with balloon angioplasty and stent placement with successful retrieval of the umbrella portion of the AED using the balloon bridge technique after failure of retrieval with the standard technique.

Case presentation

After successful revascularization of the asymptomatic severe carotid bulb stenosis in a 62-year-old woman, we could not pass the re-sheathing catheter over the deployed stent because of the ledge effect between the guidewire of the AED and the stent mesh. However, using the balloon bridge technique, which helped minimize the ledge effect, we could advance the guiding catheter beyond the stented segment over the partially inflated balloon. The umbrella portion of the AED could be easily retrieved through the guiding catheter without complications.

Conclusion

During CAS, the balloon bridge technique could be used to retrieve the AED after failure of retrieval with the standard techniques.

SUBMITTER: Kurniawan RG 

PROVIDER: S-EPMC10370601 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Successful Retrieval of the Distal Anti-embolic Device after Carotid Stenting Using the Balloon Bridge Technique.

Kurniawan Ricky G RG   Al-Abdulwahhab Abdulrahman Hamad AH   Song Yunsun Y   Kwon Boseong B   Suh Dae Chul DC   Lee Deok Hee DH  

Journal of neuroendovascular therapy 20200525 8


<h4>Background</h4>During carotid artery stenting (CAS), retrieval of the distal umbrella portion of the anti-embolic device (AED) could be difficult. Herein, we report a case of left CAS managed with balloon angioplasty and stent placement with successful retrieval of the umbrella portion of the AED using the balloon bridge technique after failure of retrieval with the standard technique.<h4>Case presentation</h4>After successful revascularization of the asymptomatic severe carotid bulb stenosi  ...[more]

Similar Datasets

| S-EPMC11220293 | biostudies-literature
| S-EPMC6581497 | biostudies-literature
| S-EPMC10092178 | biostudies-literature
| S-EPMC8917359 | biostudies-literature
| S-EPMC11640039 | biostudies-literature
| S-EPMC11576362 | biostudies-literature
| S-EPMC9746760 | biostudies-literature
| S-EPMC6624124 | biostudies-literature