An In Vitro Blood Flow Loop System for Evaluating the Thrombogenicity of Medical Devices and Biomaterials.


ABSTRACT: A reliable in vitro dynamic test method to evaluate device thrombogenicity is very important for the improvement of the design and safety of blood-contacting medical devices, while reducing the use of animal studies. In this study, a recirculating flow loop system was developed for thrombogenicity testing, using donor sheep blood anticoagulated with Anticoagulant Citrate Dextrose Solution A (ACDA) and used within 24-36 hr postdraw. Immediately before testing, the blood was recalcified and heparinized to a donor-specific target concentration. The heparinization level was based on a static pretest, in which latex tubes were incubated at room temperature for 30 min in blood with a series of heparin concentrations and evaluated for thrombus deposition. For dynamic testing, blood was recirculat