Rotational stability and refractive outcomes of the DFT/DATx15 toric, extended depth of focus intraocular lens.


ABSTRACT:

Purpose

To quantitatively assess postoperative rotational stability and visual acuity with the DFT/DATx15 extended depth of focus (EDOF) toric intraocular lens (IOL).

Methods

In this prospective case series, thirty-five patients with a calculated IOL power between + 15.0 D and + 25.0 D, corneal astigmatism between 0.75 D and 2.25 D, and no significant ocular pathology underwent cataract surgery. Primary outcome was rotational stability of the IOL at 1 month post-operatively. Secondary outcomes included residual refractive astigmatism, absolute residual astigmatism prediction error, and monocular distance and intermediate visual acuities.

Results

Mean absolute postoperative IOL rotation was 1.1 ± 0.2 degrees, with no rotation of more than 3 degrees at the final visit. Monocular mean best spectacle-corrected distance visual acuity (BSCDVA) improved from logMAR 0.27 ± 0.030 to 0.078 ± 0.017 (P < .001). Monocular uncorrected distance visual acuity (UCDVA) improved from 0.93 ± 0.096 to 0.18 ± 0.022 (P < .001). Best spectacle-corrected intermediate visual acuity (DSCIVA) was 0.17 ± 0.025, and uncorrected intermediate visual acuity (UCIVA) was 0.27 ± 0.040. Residual regular astigmatic refractive error was 0.21 ± 0.047 D.

Conclusions

The toric DFT/DATx15 EDOF lens showed excellent rotational stability and effective and predictable correction of astigmatism. Its refractive outcomes and safety profile were similar to those identified in prior studies of the non-toric DFT/DAT015 EDOF IOL. A small difference in monocular BSCDVA, of uncertain clinical significance, was found when comparing these outcomes with prior DFT/DAT015 data. The trial was retrospectively registered on November 5, 2021 (TRN ​​NCT05119127).

SUBMITTER: Barber KM 

PROVIDER: S-EPMC10371906 | biostudies-literature | 2023 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Rotational stability and refractive outcomes of the DFT/DATx15 toric, extended depth of focus intraocular lens.

Barber Kevin M KM   O'Connor Sara S   Mackinder Philip P   Chih Andreea A   Jones Brian B  

International ophthalmology 20230310 8


<h4>Purpose</h4>To quantitatively assess postoperative rotational stability and visual acuity with the DFT/DATx15 extended depth of focus (EDOF) toric intraocular lens (IOL).<h4>Methods</h4>In this prospective case series, thirty-five patients with a calculated IOL power between + 15.0 D and + 25.0 D, corneal astigmatism between 0.75 D and 2.25 D, and no significant ocular pathology underwent cataract surgery. Primary outcome was rotational stability of the IOL at 1 month post-operatively. Secon  ...[more]

Similar Datasets

| S-EPMC7457597 | biostudies-literature
| S-EPMC7171658 | biostudies-literature
| S-EPMC9905548 | biostudies-literature
| S-EPMC8555820 | biostudies-literature
| S-EPMC8761297 | biostudies-literature
| S-EPMC7002424 | biostudies-literature
| S-EPMC9753564 | biostudies-literature
| S-EPMC8302686 | biostudies-literature
| S-EPMC7249429 | biostudies-literature
| S-EPMC7326203 | biostudies-literature