Ontology highlight
ABSTRACT: Background
High-power, short-duration (HPSD) radiofrequency ablation (RFA) reduces procedure time; however, safety and efficacy thresholds vary with catheter design.Objective
The study sought to determine optimal HPSD ablation conditions with a novel flexible-tipped, contact force-sensing RFA catheter.Methods
RFA lesions were created in thigh muscle (16 swine) over a range of conditions (51-82 W, 2-40 g, 8-40 mL/min irrigation). An intracardiac study was performed (12 swine) to characterize steam pop thresholds. Lesions were created in a second intracardiac study (14 swine, n = 290 pulmonary vein isolation [PVI] lesions) with combinations of radiofrequency power, duration, and contact force. PVI was tested, animals were sacrificed, and lesions were measured.Resu
SUBMITTER: Piccini JP