Parameter based 4D dose calculations for proton therapy.
Ontology highlight
ABSTRACT: Background and purpose Retrospective log file-based analysis provides the actual dose delivered based on the patient's breathing and the daily beam-delivery dynamics. To predict the motion sensitivity of the treatment plan on a patient-specific basis before treatment start a prospective tool is required. Such a parameter-based tool has been investigated with the aim to be used in clinical routine. Materials and Methods 4D dose calculations (4DDC) were performed for seven cancer patients with small breathing motion treated with scanned pulsed proton beams. Validation of the parameter-based 4DDC (p-4DDC) method was performed with an anthropomorphic phantom and patient data employing measurements and a log file-based 4DDC tool. The dose volume histogram parameters (Dx%) were inv
SUBMITTER: Lebbink F
PROVIDER: S-EPMC10374597 | biostudies-literature | 2023 Jul