Unknown

Dataset Information

Validation of robust radiobiological optimization algorithms based on the mixed beam model for intensity-modulated carbon-ion therapy.


ABSTRACT: Currently, treatment planning systems (TPSs) that can compute the intensities of intensity-modulated carbon-ion therapy (IMCT) using scanned carbon-ion beams are limited. In the present study, the computational efficacy of the newly designed IMCT algorithms was analyzed for the first time based on the mixed beam model with respect to the physical and biological doses; moreover, the validity and effectiveness of the robust radiobiological optimization were verified. A dose calculation engine was independently generated to validate a clinical dose determined in the TPS. A biological assay was performed using the HSGc-C5 cell line to validate the calculated surviving fraction (SF). Both spot control (SC) and voxel-wise worst-case scenario (WC) algorithms were employed for robust radiobiologic

SUBMITTER: Yagi M 

PROVIDER: S-EPMC10381094 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature