Simultaneous Control of Venous Reservoir Level and Arterial Flow Rate in Cardiopulmonary Bypass With a Centrifugal Pump.


ABSTRACT: Cardiopulmonary bypass (CPB) is an indispensable technique in cardiac surgery, providing the ability to temporarily replace cardiopulmonary function and create a bloodless surgical field. Traditionally, the operation of CPB systems has depended on the expertise and experience of skilled perfusionists. In particular, simultaneously controlling the arterial and venous occluders is difficult because the blood flow rate and reservoir level both change, and failure may put the patient's life at risk. This study proposes an automatic control system with a two-degree-of-freedom model matching controller nested in an I-PD feedback controller to simultaneously regulate the blood flow rate and reservoir level. CPB operations were performed using glycerin and bovine blood as perfusate to simulate flo