Accuracy of cuff blood pressure and systolic blood pressure amplification.
Ontology highlight
ABSTRACT: Automated cuff measured blood pressure (BP) is the global standard used for diagnosing hypertension, but there are concerns regarding the accuracy of the method. Individual variability in systolic BP (SBP) amplification from central (aorta) to peripheral (brachial) arteries could be related to the accuracy of cuff BP, but this has never been determined and was the aim of this study. Automated cuff BP and invasive brachial BP were recorded in 795 participants (74% male, aged 64 ± 11 years) receiving coronary angiography at five independent research sites (using seven different automated cuff BP devices). SBP amplification was recorded invasively by catheter and defined as brachial SBP minus aortic SBP. Compared with invasive brachial SBP, cuff SBP was significantly underestimated (130 ± 18