ABSTRACT: Objective
HAART 300 300 (BioStable Science and Engineering, Inc) aortic annuloplasty rings restore physiologic annular geometry during aortic valve repair. Transcatheter valve-in-ring implantation is appealing for recurrent valve dysfunction but may necessitate balloon fracture of downsized annuloplasty rings. We characterized the feasibility of ring fracture and changes in ring geometry preceding fracture.Methods
The 19-mm, 21-mm, and 23-mm HAART 300 annuloplasty rings were obtained, and 23-mm, 24-mm, 25-mm, and 26-mm valvuloplasty balloons were obtained. Under continuous fluoroscopy and video recording, a 23-mm balloon was inflated within a 19-mm ring at 1 atm/s until ring fracture or balloon failure occurred. If balloon failure occurred, experiments were sequentially rep