Real-world agreement of same-visit Tono-Pen vs Goldmann applanation intraocular pressure measurements using electronic health records.


ABSTRACT:

Purpose

To compare intraocular pressure (IOP) obtained with Tono-Pen (TP) and Goldmann applanation (GAT) using large-scale electronic health records (EHR).

Design

Retrospective cohort study.

Methods

A single pair of eligible TP/GAT IOP readings was randomly selected from the EHR for each ophthalmology patient at an academic ophthalmology center (2013-2022), yielding 4550 eligible measurements. We used Bland-Altman analysis to describe agreement between TP/GAT IOP differences and mean IOP measurements. We also used multivariable logistic regression to identify factors associated with different IOP readings in the same eye, including demographics, glaucoma diagnosis, and central corneal thickness (CCT). Primary outcome metrics were discrepant measurements between TP and