Potential of GC-Combustion-MS as a Powerful and Versatile Nitrogen-Selective Detector in Gas Chromatography.


ABSTRACT: Here, we show the potential and applicability of the novel GC-combustion-MS approach as a nitrogen-selective GC detector. Operating requirements to achieve reproducible and compound-independent formation of volatile NO species as a selective N-signal during the combustion step are described. Specifically, high temperatures (≥1000 °C) and post-column O2 flows (0.4 mL min-1 of 0.3% O2 in He) turned out to be necessary when using a vertical oven without makeup flow (prototype #1). In contrast, the use of a horizontal oven with 1.7 mL min-1 He as an additional makeup flow (prototype #2) required milder conditions (850 °C and 0.2 mL min-1). A detection limit of 0.02 pg of N injected was achieved, which is by far the lowest ever reported for