Disagreement of Radial Peripapillary Capillary Density Among Four Optical Coherence Tomography Angiography Devices.


ABSTRACT:

Purpose

This prospective study evaluated the agreement among four optical coherence tomography angiography (OCTA) devices in the assessment of radial peripapillary capillary (RPC) density.

Methods

The study included 48 eyes of 48 subjects (14 healthy, 19 glaucomatous, and 15 non-glaucomatous optic neuropathy). Each participant was scanned using four OCTA devices in a random sequence: RTVue XR Avanti (RTVue), DRI OCT Triton (Triton), Revo NX 130 (Revo), and PLEX Elite 9000 (PlexE). All 6 × 6-mm grayscale OCTA images from each device were analyzed for RPC density using a customized algorithm. Agreement between each pair of devices was assessed using intraclass correlation coefficients (ICCs) and Bland-Altman plots.

Results

There was a poor correlation between devices in