Twelve-Month Clinical and Histopathological Performance of a Novel Synthetic Cornea Device in Rabbit Model.


ABSTRACT:

Purpose

To report the biological stability and postoperative outcomes of a second-generation, single-piece, flexible synthetic cornea in a rabbit model.

Methods

Device materials and design were amended to enhance biointegration. Optic skirt design devices were made from compact perfluoroalkoxy alkane with porous expanded polytetrafluoroethylene ingrowth surface overlying the skirt and optic wall. Sixteen devices were implanted into intrastromal pocket in rabbit eyes. Rabbits were randomly assigned to 6- and 12-month follow-up cohorts (n = 8 in each) postoperatively. Monthly examinations and optical coherence tomography assessed cornea-device integration, iridocorneal angle, optic nerve, and retina.

Results

There were no intraoperative complications. All devices were i