Unknown

Dataset Information

Quick versus Quantitative: Evaluation of Two Commercial Real-Time PCR Assays for the Detection of Pneumocystis jirovecii from Bronchoalveolar Lavage Fluids.


ABSTRACT: Two commercial real-time PCR assays for the detection of Pneumocystis jirovecii were compared, the quantitative RealStar P. jirovecii assay and the qualitative DiaSorin P. jirovecii assay, the latter of which can be used without nucleic acid extraction. Archived bronchoalveolar lavage (BAL) specimens (n = 66), previously tested by molecular methods, were tested by both assays, and the results were compared to the respective original result. The RealStar P. jirovecii assay demonstrated good positive percent agreement (PPA) (90% [95% confidence interval (CI), 72 to 97%]; 27/30) and negative percent agreement (NPA) (100% [95% CI, 88 to 100%]; 36/36) with the reference method. The DiaSorin P. jirovecii assay concordantly detected P. jirovecii in 19 of 24 positive BAL samples (PPA = 73%

SUBMITTER: Belanger CR 

PROVIDER: S-EPMC10434167 | biostudies-literature | 2023 Aug

REPOSITORIES: biostudies-literature