ABSTRACT: Background
Mild secretion defects are the most frequent and challenging blood platelet disorders to diagnose. Most δ-granule secretion tests lack validation, are not quantitative, or have unreliable response to weak platelet agonists.Objectives
To compare platelet serotonin secretion by HPLC-electrochemical detection technique (HPLC-ECD) with the reference isotopic test (3H-5-HT), evaluating its performance in clinical laboratories.Methods
The assay validation followed STARD-2015 recommendations. HPLC-ECD measured the nonsecreted serotonin remaining in platelet pellets after aggregation, comparing it with the reference 3H-5-HT assay. We studied subjects with inherited and aspirin-induced blood platelet disorders and assessed the HPLC-ECD operat